Osteotome Failure

Technical Challenge

An osteome unexpectedly failed during a plastic surgery operation. Veryst was hired to explain the failure.

Veryst Medical Solution

Scanning electron microscopy of the fracture surface showed intergranular corrosion at the cutting surface. A review of the hospital’s sterilization procedures revealed the use of an alkaline cleaner specifically contraindicated for the alloy used in the osteotome, as its use could weaken the material. Identification of cleaning issue enabled the prevention of similar failures in other osteomes of the same alloy.




Example Project: Osteotomes
Broken osteotome end

Example Project: Osteotomes
SEM image of broken osteotome end